6.12 RATE Group

The command in this group deals with the data update interval. You can make the same settings and queries that you
can by pressing SETUP on the front panel and using the u.rAtE menu.
:RATE
Function
Sets or queries the data update interval.
Syntax
:RATE {< Time >|AUTO}
:RATE?
< Time > = 100 , 250 , 500(ms) , 1 , 2 , 5 , 10 ,
20(s)
Example
:RATE 250MS
:RATE? -> :RATE 250.0E-03
:RATE:AUTO?
Function
Queries all applicable settings for when the data
update interval is set to Auto.
Syntax
:RATE:AUTO?
:RATE:AUTO:TIMeout
Function
Sets or queries the timeout for when the data
update interval is set to Auto.
Syntax
:RATE:AUTO:TIMeout {<NRf>}
:RATE:AUTO:TIMeout?
<NRf> = 1 , 5 , 10 , 20(s)
Example
:RATE:AUTO:TIMEOUT 1
:RATE:AUTO:TIMEOUT?
-> :RATE:AUTO:TIMEOUT 1
:RATE:AUTO:SYNChronize
Function
Sets or queries the synchronization source for
when the data update interval is set to Auto.
Syntax
:RATE:AUTO:SYNChronize {U<x>|I<x>}
:RATE:AUTO:SYNChronize?
<x> = 1 to 3 (element)
Example
:RATE:AUTO:SYNCHRONIZE U1
:RATE:AUTO:SYNCHRONIZE?
-> :RATE:AUTO:SYNCHRONIZE U1