YOKOGAWA WT310EH User Manual page 53

Digital power meter communication interface
Hide thumbs Also See for WT310EH:
Table of Contents

Advertisement

[:INPut]:VOLTage:POJump
Function
Sets or queries the jump destination range that is
used when a voltage peak over-range occurs.
Syntax
[:INPut]:VOLTage:POJump
{OFF|< Voltage >}
[:INPut]:VOLTage:POJump?
OFF = No jump destination voltage range
<Voltage> = See ( :INPut:VOLTage:RANGe ).
Example
:INPUT:VOLTAGE:POJUMP 600V
:INPUT:VOLTAGE:POJUMP?
-> :INPUT:VOLTAGE:POJUMP 600.0E+00
[:INPut]:CURRent?
Function
Queries all electric current measurement settings.
Syntax
[:INPut]:CURRent?
[:INPut]:CURRent:RANGe
Function
Sets or queries the current range.
Syntax
[:INPut]:CURRent:RANGe
{< Current >|(EXTernal,< Voltage >)}
[:INPut]:CURRent:RANGe?
• For direct current input
• When the crest factor is set to 3
<Current> = 5, 10, 20, 50, 100, 200, 500(mA),
1, 2, 5, 10, 20(A) (WT310E)
<Current> = 1, 2, 5, 10, 20, 40(A) (WT310EH)
<Current> = 0.5, 1, 2, 5, 10, 20(A) (WT332E,
WT333E)
• When the crest factor is set to 6 or 6A
<Current> = 2.5, 5, 10, 25, 50, 100, 250(mA),
0.5, 1, 2.5, 5, 10(A) (WT310E)
<Current> = 0.5, 1, 2.5, 5, 10, 20(A) (WT310EH)
<Current> = 0.25, 0.5, 1, 2.5, 5, 10(A) (WT332E,
WT333E)
• For external current sensor input
• When the crest factor is set to 3
<Voltage> = 2.5, 5, 10(V) (/EX1)
<Voltage> = 50, 100, 200, 500(mV), 1, 2(V)
(/EX2)
• When the crest factor is set to 6 or 6A
<Voltage> = 1.25, 2.5, 5(V) (/EX1)
<Voltage> = 25, 50, 100, 250(mV), 0.5, 1(V)
(/EX2)
Example
:INPUT:CURRENT:RANGE 20A
:INPUT:CURRENT:RANGE?
-> :INPUT:CURRENT:RANGE 20.0E+00
:INPUT:CURRENT:RANGE EXTERNAL,10V
:INPUT:CURRENT:RANGE?
-> :INPUT:CURRENT:RANGE
EXTERNAL,10.0E+00
Description EXTernal and <Voltage> can only be selected
on models with the external current sensor input
(/EX1 or /EX2) option.
IM WT310E-17EN
[:INPut]:CURRent:AUTO
Sets or queries the current auto range on/off state.
Function
Syntax
[:INPut]:CURRent:AUTO {<Boolean>}
[:INPut]:CURRent:AUTO?
Example
:INPUT:CURRENT:AUTO ON
:INPUT:CURRENT:AUTO?
-> :INPUT:CURRENT:AUTO 1
[:INPut]:CURRent:CONFig
Function
Sets or queries the valid current range.
Syntax
[:INPut]:CURRent:CONFig {ALL|< Current >
[,< Current >][,< Current >]...}
[:INPut]:CURRent:CONFig?
ALL = All ranges are valid.
<Current> = See ( :INPut:CURRent:RANGe ).
Example
:INPUT:CURRENT:CONFIG ALL
:INPUT:CURRENT:CONFIG?
-> :INPUT:CURRENT:CONFIG ALL
:INPUT:CURRENT:CONFIG 20,5,1
:INPUT:CURRENT:CONFIG?
-> :INPUT:CURRENT:CONFIG
20.0E+00,5.0E+00,1.0E+00
Description In the parameters, list the current ranges that you
want to enable. To enable all the ranges, specify
the parameter "ALL."
[:INPut]:CURRent:POJump
Function
Sets or queries the jump destination range that is
used when a current peak over-range occurs.
Syntax
[:INPut]:CURRent:POJump
{OFF|< Current >}
[:INPut]:CURRent:POJump?
OFF = No jump destination current range
<Current> = See ( :INPut:CURRent:RANGe ).
Example
:INPUT:CURRENT:POJUMP 20A
:INPUT:CURRENT:POJUMP?
-> :INPUT:CURRENT:POJUMP 20.0E+00
[:INPut]:CURRent:EXTSensor:CONFig
Function
Sets or queries the valid external current sensor
range.
Syntax
[:INPut]:CURRent:EXTSensor:CONFig
{ALL|< Voltage >[,< Voltage >]
[,< Voltage >]...}
[:INPut]:CURRent:EXTSensor:CONFig?
ALL = All ranges are valid.
<Voltage> = See ( :INPut:CURRent:RANGe ).
Example
:INPUT:CURRENT:EXTSENSOR:CONFIG ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG? ->
:INPUT:CURRENT:EXTSENSOR:CONFIG ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG
2,0.5,0.1
:INPUT:CURRENT:EXTSENSOR:CONFIG?
-> :INPUT:CURRENT:EXTSENSOR:CONFIG
2.00E+00,500.0E-03,100.0E-03
6.7 INPut Group
1
2
3
4
5
6
7
8
9
App
Index
6-13