:NUMeric:LIST?
Function
Queries all harmonic measurement numeric list
data output settings.
Syntax
:NUMeric:LIST?
Description • This is only valid on models with the harmonic
measurement (/G5) option.
• The number of numeric list data items output
by :NUMeric:LIST:ITEM<x> is determined by
:NUMeric:LIST:NUMber.
:NUMeric:LIST:VALue?
Function
Queries the harmonic measurement numeric list
data.
Syntax
:NUMeric:LIST:VALue? {<NRf>}
<NRf> = 1 to 32 (item number)
• When <NRf> is specified
Example
:NUMERIC:LIST:VALUE? 1
-> 103.58E+00,NAN,103.53E+00
,0.09E+00,2.07E+00,0.04E+00,..
(omitted) ..,0.01E+00,0.01E+00
(up to 52 data values)
• When <NRf> is omitted (when :NUMeric
:LIST:NUMber is set to 5)
:NUMERIC:LIST:VALUE?
-> 103.58E+00,NAN,103.53E+00
,0.09E+00,2.07E+00,0.04E+00,..
(omitted) ..,0.00E+00,0.00E+00
(up to 52*5 = 260 data values)
• When :NUMeric:FORMat is set to {FLOat}
:NUMERIC:LIST:VALUE?
-> #N (N-digit byte number)(data byte
sequence)
Description • This is only valid on models with the harmonic
measurement (/G5) option.
• A single numeric list data item consists
of up to 52 items of numeric data in the
following order: TOTal, DC, 1st harmonic, ...,
:NUMeric:LIST:ORDer.
• If <NRf> is specified, only the numeric list data
of the specified item number is output (up to 52
items of data).
• If <NRf> is omitted, the numeric list data of
item numbers from 1 to :NUMeric:LIST:NUMber
is output in order (up to 52 times the number
specified by :NUMeric:LIST:NUMber).
• For the formats of each individual numeric data
output item, see "Numeric Data Formats" at the
end of this group of commands (page 6-25).
• This instrument does not measure data for the
DC component. It is always "NAN."
IM WT310E-17EN
:NUMeric:LIST:NUMber
Function
Sets or queries the number of numeric
list data items that are transmitted by
:NUMeric:LIST:VALue?.
Syntax
:NUMeric:LIST:NUMber {<NRf>|ALL}
:NUMeric:LIST:NUMber?
<NRf> = 1 to 32(ALL)
Example
:NUMERIC:LIST:NUMBER 5
:NUMERIC:LIST:NUMBER
-> :NUMERIC:LIST:NUMBER 5
Description • This is only valid on models with the harmonic
measurement (/G5) option.
• If the parameter is omitted from the
:NUMeric:LIST:VALue? command, the numeric
list data items from 1 to the specified value are
output in order.
• By default, the number of numeric list data
items is set to 1.
:NUMeric:LIST:ORDer
Function
Sets or queries the maximum output harmonic
order of the harmonic measurement numeric list
data.
Syntax
:NUMeric:LIST:ORDer {<NRf>|ALL}
:NUMeric:LIST:ORDer?
<NRf> = 1 to 50(ALL)
Example
:NUMERIC:LIST:ORDER 50
:NUMERIC:LIST:ORDER?
-> :NUMERIC:LIST:ORDER 50
Description This is only valid on models with the harmonic
measurement (/G5) option.
:NUMeric:LIST:SELect
Function
Sets or queries the output components of the
harmonic measurement numeric list data.
Syntax
:NUMeric:LIST:SELect {EVEN|ODD|ALL}
:NUMeric:LIST:SELect?
Example
:NUMERIC:LIST:SELECT ALL
:NUMERIC:LIST:SELECT?
-> :NUMERIC:LIST:SELECT ALL
Description • This is only valid on models with the harmonic
measurement (/G5) option.
• The available options are explained below.
EVEN = Outputs the components of TOTal,
DC, and even-order harmonics
ODD = Outputs the components of TOTal, DC,
and odd-order harmonics
ALL = Outputs all components
6.11 NUMeric Group
App
Index
6-21
1
2
3
4
5
6
7
8
9