6.4

DISPlay group

The commands in this group deal with the display.
You can make the same settings and queries that you can by using keys such as the FUNCTION and ELEMENT
keys on the front panel.
:DISPlay?
Function
Queries all display settings.
Syntax
:DISPlay?
:DISPlay:NORMal?
Function
Queries all normal measurement data display
settings.
Syntax
:DISPlay:NORMal?
:DISPlay[:NORMal]:ITEM<x>
Function
Sets or queries a normal measurement data
display item.
Syntax
:DISPlay[:NORMal]:ITEM<x> <Function>
[,<Element>]}
:DISPlay[:NORMal]:ITEM<x>?
<x> = 1 to 4 (display)
Function of display A ( <x>=1 )
<Function> = {U|I|P|S|Q|TIME}
Function of display B ( <x>=2 )
<Function> = {U|I|P|LAMBda|PHI}
Function of display C ( <x>=3 )
<Function> = {U|I|P|UPPeak|UMPeak|
IPPeak|IMPeak|PPPeak|PMPeak|WH|
WHP|WHM|AH|AHP|AHM|MATH}
Function of display D ( <x>=4 )
<Function> = {U|I|P|LAMBda|FU|FI|
UTHD|ITHD}
<Element> = {<NRf>|SIGMa}
(<NRf> = 1 to 3)
Example
:DISPLAY:NORMAL:ITEM1 U,1
:DISPLAY:NORMAL:ITEM1?
-> :DISPLAY:NORMAL:ITEM1 U,1
Description • For details on <Function>, see "Numeric Data
Display Functions" on the next page.
• If <Element> is omitted, the element is set to 1.
• For {TIME|MATH}, <Element> does not need
to be specified. In responses, <Element> is
omitted.
• {UTHD|ITHD} can be selected only on models
with the harmonic measurement (/G5) option.
6-8
:DISPlay:HARMonics?
Function
Queries all harmonic measurement data display
settings.
Syntax
:DISPlay:HARMonics?
:DISPlay:HARMonics:ITEM<x>
Function
Sets or queries a harmonic measurement data
display item.
Syntax
:DISPlay:HARMonics:ITEM<x>
{<Function>[,<Element>]}
:DISPlay:HARMonics:ITEM<x>?
<x> = 1 to 4 (display)
Function of display A ( <x>=1 )
<Function> = {U|I|P|ORDer}
Function of display B ( <x>=2 )
<Function> = {U|I|P|UHDF|IHDF|PHDF|
PHIU|PHII}
Function of display C ( <x>=3 )
<Function> = {U|I|P}
Function of display D ( <x>=4 )
<Function> = {U|I|P|LAMBda|FU|FI|
UTHD|ITHD}
<Element> = {<NRf>} (<NRf> = 1 to 3)
Example
:DISPLAY:HARMONICS:ITEM2 I,1
:DISPLAY:HARMONICS:ITEM2?
-> :DISPLAY:HARMONICS:ITEM2 I,1
Description • For details on <Function>, see "Numeric Data
Display Functions" on the next page.
• If <Element> is omitted, the element is set to 1.
• For {ORDer}, <Element> does not need to be
specified. In responses, <Element> is omitted.
IM WT310E-17EN