6.17 Common Command Group
*TST?
Executes a self-test and queries the result.
Function
Syntax
*TST?
Example
*TST? -> 0
Description • The self-test consists of tests of each kind of
internal memory.
• This command returns 0 if the self-test is
successful and 1 if it is not.
• It takes approximately 6 seconds for the test to
complete. When receiving a response from this
instrument, set the timeout to a relatively large
value.
*WAI
Function
Holds the execution of the subsequent command
until the completion of the specified overlap
command.
Syntax
*WAI
Example
*WAI
Description This instrument does not have overlap
commands. This command will be ignored.