Software-Based Option Installation

You can install or uninstall a software-based option on the ultrasound
system using an MO disk and the MO drive. One key disk is required for
one option per system. You cannot concurrently install a key disk option on
multiple systems. If subsequently uninstalled, the option key disk can be
installed on a different ultrasound system.
Observe the following precautions when using a key disk:
Do not power on the ultrasound system with a key disk inserted in the
MO drive.
Do not remove the key disk during option installation or uninstallation.
To install a software-based option:
1. Insert the media containing the option into the MO drive on the left of
the system.
2. Press the F4 key on the keyboard.
The system displays the Preset Main Menu.
3. Select Installation from Key Disk.
4. Select the Install button.
The system displays a confirmation dialog box.
5. Select the OK button.
The option is installed. When the installation is complete, the system
displays a new message box.
6. Select the OK button.
The option is listed on the Installed Option screen.
7. Press the eject button near the MO drive and remove the key disk.
[ 1 ]