MAMMOMAT 1000/3000 Nova - Opdimar
Service
Service Instructions
ASW V3.1
Register 5
Print No.: SPB7-230.061.11.02.02
Replaces: SPB7-230.061.11.01.02
© Siemens AG 2003
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model
_or_ design,_are_
reserved.
English
Doc. Gen. Date: 07.03
(66 31 951)

Advertisement

Table of Contents
loading

Summary of Contents for Siemens MAMMOMAT Nova Opdima 1000

  • Page 1 MAMMOMAT 1000/3000 Nova - Opdimar Service Service Instructions ASW V3.1 © Siemens AG 2003 The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights,...
  • Page 2: Document Revision Level

    Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho- rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
  • Page 3: Table Of Contents

    Enabling the DICOM option ......5 - 15 Instruction for obtaining the systems host ID and hostname ... 5 - 15 Siemens AG Register 5 SPB7-230.061.11...
  • Page 4 Biopsy unit not responding ......9 - 2 Opdimar Register 5 SPB7-230.061.11 Page 4 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24...
  • Page 5 Biopsy controller cable duct ......10 - 6 11 ______Changes to previous version ____________________________________ 11 - 1 12 ______Appendix 1 __________________________________________________ A1 - 1 Siemens AG Register 5 SPB7-230.061.11...
  • Page 6 Biopsy controller cable duct ......14 - 6 Opdimar Register 5 SPB7-230.061.11 Page 6 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24...
  • Page 7: General

    OPDIMAr Solaris 8 Installation CD v1.0, part No. 66 33 049 (included in the Opdimar delivery) • OPDIMAr Installation CD, version 3.1, part No. 66 33 700 (included in the Opdimar delivery) Siemens AG Register 5 SPB7-230.061.11 Page 1 of 2 Opdimar Medical Solutions Rev.
  • Page 8: Meters And Appliances Required

    Sun Blade 150 Service manuals). Tools required • Standard service tools • PROM extractor for PLCC 32 Opdimar Register 5 SPB7-230.061.11 Page 2 of 2 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 9: Block Diagram

    D110 CCD Printer camera board (option) 5V_COM control logic C_CL CCD_R A_CL SRC_CL CCD sandwich = Galvanic separation Fig. 1 Block diagram Siemens AG Register 5 SPB7-230.061.11 Page 1 of 4 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 10: Biopsy Controller

    Selection of RS-232 communication to MAMMOMAT printer/service PC or workstation (SELECT_WS) • Supply voltages (e.g. 5V_COM, 30V_CC) For more information on LEDs, see MAMMOMAT 1000/3000/3000 Nova - Opdimar Wiring Diagram. Opdimar Register 5 SPB7-230.061.11 Page 2 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 11: Workstation

    Stereo examination Spot examination Database Service mode mode mode mode Data storage Help (n.a.) Error handling Fig. 2 Application software logic flow Siemens AG Register 5 SPB7-230.061.11 Page 3 of 4 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 12: Ccd Camera Communication

    Three potentiometers monitors the actual position of the needle tip. Opdimar Register 5 SPB7-230.061.11 Page 4 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 13 The camera shall be used within 10230i C. The camera shall be transported or stored within 0240i C. Siemens AG Register 5 SPB7-230.061.11 Page 1 of 2...
  • Page 14 3 - 2 Protective measures for CCD camera This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 2 of 2 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 15 Use ESD-equipment, ground prior to making contact and place the components on a conductive surface. The boards contain electrostatic highly sensitive components requiring particular care in their handling. Risk of damaging components. Siemens AG Register 5 SPB7-230.061.11 Page 1 of 2 Opdimar Medical Solutions Rev.
  • Page 16 4 - 2 Protective measures This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 2 of 2 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 17: General

    Advanced service Service users can additionally access advanced service functions. To get access you have to log in with the user name “service”. The password is obtained from Siemens Head Quarter Support Center. The following advanced service functions are included: •...
  • Page 18: Selection Of Mode

    Select desired service function from the service dialog. Service Biopsy Calibration Needle Setup Network Setup User Administration Advanced Service Change Password Backup Close Help Fig. 2 Service dialog Opdimar Register 5 SPB7-230.061.11 Page 2 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 19: Calibration Of The Biopsy Unit

    With the targets facing the patient side, the phantom fits in the opening of the stereo compression plate, see Fig. 4. Target 2 shall be positioned 25228 mm in negative x-axis direction and 12 mm from the object table side. Siemens AG Register 5 SPB7-230.061.11...
  • Page 20 Acquire and release a stereo pair of images. Set the exposure parameters to 25 kV and 28 mAs in manual mode. Opdimar Register 5 SPB7-230.061.11 Page 4 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 21 Fig. 6 Information message Perform a final check according to Verifying the calibration of the biopsy unit on Page 10 - 1 to make sure that the biopsy unit works properly. Siemens AG Register 5 SPB7-230.061.11 Page 5 of 30...
  • Page 22: Setting Up Needles

    L1 shall be 30 to 175 mm when selecting fine needle, L2 is not applicable for fine needle. When selecting core, L1 shall be set to <170 mm. Opdimar Register 5 SPB7-230.061.11 Page 6 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 23: Editing Needle Values

    When using core gun, the system safety margin to avoid hitting the biopsy table is 5 mm. Editing needle values Select a needle from the needle setup dialog. Press Edit..Siemens AG Register 5 SPB7-230.061.11 Page 7 of 30 Opdimar Medical Solutions Rev.
  • Page 24: Backup Of Temporary Storage Media

    This backup might take a long time and require several MO disks NOTICE since all folders on hard disk will be copied to backup MO. Opdimar Register 5 SPB7-230.061.11 Page 8 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 25: Use Of Advanced Service Functions

    General To get access to the advanced service menu you have to log in with the user name “service”. The password is obtained from Siemens Uptime Service Center (or from Headquarter Support Center). The advanced service is carried out by using a number of text dialogs. Selection of an item in a dialog can be done in three different ways: •...
  • Page 26 Log Administration Function for log settings. View Log Function for inspection of logs. Close Function for leaving the advanced service mode. Opdimar Register 5 SPB7-230.061.11 Page 10 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 27: Country Settings

    Set the values for Language, Time Zone, ID Pattern, Date Pattern, Service Center, Institution Name and Department Name in each respective fields. Select Apply and press Enter to execute the changes. Siemens AG Register 5 SPB7-230.061.11 Page 11 of 30...
  • Page 28: Test Of Units

    Biopsy controller - test if there is a connection and if the power is on • Database - reading and writing in the database Opdimar Register 5 SPB7-230.061.11 Page 12 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 29: Software Upgrade

    The hostname of the original workstation can be restored to a new workstation. Restore hostname from most recently used local/backup MO disk. Update eeprom Not applicable. Miscellaneous Database and MO disk utilities. Siemens AG Register 5 SPB7-230.061.11 Page 13 of 30 Opdimar Medical Solutions Rev. 02 07.03...
  • Page 30: New Software Version

    NEW SOFTWARE VERSION Current Release: opdima.3.0_6633064 CD-ROM Release: opdima.3.1_6633700 Action: Close Close/Help/Install Fig. 15 New software version, advanced service dialog Select Install. Opdimar Register 5 SPB7-230.061.11 Page 14 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 31: Enabling The Dicom Option

    The DICOM licence is written on the invoice and the dispatch note. If not, please contact Siemens AG Medical Solutions in Erlangen, Headquarter Support Center of Special Prod- ucts Division (Dept. CS HSC 24) via phone (+49/ (0) 9191 18 8080.1.6.3) for further infor- mation.
  • Page 32 Select y to reboot. Configure DICOM nodes according to Network Setup in Supplement to the Instructions for Use MAMMOMAT 3000 - Opdimar. Opdimar Register 5 SPB7-230.061.11 Page 16 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 33: Network Settings

    Some alive routers do not reply to pingtest, check with network administrator. To test if a host is alive or not, write the IP address temporarily in the field for Default router and perform a pingtest. Siemens AG Register 5 SPB7-230.061.11 Page 17 of 30...
  • Page 34: Miscellaneous

    Action: Close Close/Help/Run Fig. 19 Check database, advanced service dialog Select the database to check, MO Disk and/or Local Database. Select Run. Opdimar Register 5 SPB7-230.061.11 Page 18 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 35 Format a new disk or erase contents of a formerly used disk. WARNING! Use with extreme caution. Erases all content on MO disk. WARNING! When using this function, It is only possible to use a 9.1 GB MO disk. Siemens AG Register 5 SPB7-230.061.11 Page 19 of 30 Opdimar Medical Solutions Rev.
  • Page 36 Select refresh rate according to the chapter Installation of monitor in the Installation and Start-Up Instructions. Select Apply. NOTICE To enable the new settings reboot the system. Opdimar Register 5 SPB7-230.061.11 Page 20 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 37: Restoring Data From Mo Disk

    MO disk num- bering can be incorrect. Please contact HSC for more information. NOTICE Set up the printer once again after software reinstallation. Printer info can not be restored. Siemens AG Register 5 SPB7-230.061.11 Page 21 of 30 Opdimar Medical Solutions Rev.
  • Page 38: Ccd Camera Calibration And Maintenance

    Grid slow speed % of max <ESC> to exit, <TAB> move to next entry field 1 Help 2 Save 10 Quit Fig. 25 Opdimar Register 5 SPB7-230.061.11 Page 22 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 39: Non Grid Table

    “Camera successfully calibrated” will appear on the screen. Perform check of Opdimar AEC and check of resolution. See Check of Opdima AEC on Page 10 - 2 and Check of resolution on Page 10 - 4. Siemens AG Register 5 SPB7-230.061.11...
  • Page 40: Disk Cache Settings

    MO disk. If a folder has been copied to a MO disk, its images will be erased from the hard disk if the cache limit is reached. Opdimar Register 5 SPB7-230.061.11 Page 24 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 41: Use Of Printer Setup

    Select Apply and press Enter. Connect your selected printer after turning off the system. If no printer is connected select none, apply and restart the sys- NOTICE tem. Siemens AG Register 5 SPB7-230.061.11 Page 25 of 30 Opdimar Medical Solutions Rev.
  • Page 42: Networked Opdima System

    Textual description, for example the location of the printer. E.g. Codonics_mammo_dept. Address: IP address for printer, e.g. 123.123.123.33. No initial 0, e.g. 10.10.10.10 is OK but 010.010.010.01 is not OK. Opdimar Register 5 SPB7-230.061.11 Page 26 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 43 Close/Help/Apply Fig. 31 Remove network printer dialog, example Previously defined printers can be removed from the list of printers. Note that the printer that is currently selected cannot be removed. Siemens AG Register 5 SPB7-230.061.11 Page 27 of 30 Opdimar Medical Solutions Rev.
  • Page 44: Log Administration

    Select Log Administration in the advanced service dialog. Select Yes/No to turn on/off each respective log. Select Apply and press Enter. Opdimar Register 5 SPB7-230.061.11 Page 28 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 45: Log Inspection

    Log for activities made in backup mode. Close Function for leaving the view log dialog. Select View Log in the advanced service dialog. Select a log from the menu in the view log dialog. Siemens AG Register 5 SPB7-230.061.11 Page 29 of 30 Opdimar Medical Solutions Rev.
  • Page 46 5 - 30 Service mode This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 30 of 30 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 47: General

    Disconnect the power supply cable (AC in) and SCSI cable from the back of the unit (Fig. 2). Unplug the SCSI terminator from the back of the unit (Fig. 2). Remove the unit. Siemens AG Register 5 SPB7-230.061.11 Page 1 of 14...
  • Page 48: Installing The Mo Unit

    F.GND (frame ground) terminal Emergency eject hole AC IN (AC power) connector EJECT button Fig. 2 MO unit, front and rear panel Opdimar Register 5 SPB7-230.061.11 Page 2 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 49: Tests And Adjustments

    Connect the power supply cable, the SCSI terminator and the SCSI cable to the unit. Tests and Adjustments Switch on the system. Ensure that the system is operating properly. Siemens AG Register 5 SPB7-230.061.11 Page 3 of 14 Opdimar Medical Solutions Rev.
  • Page 50: Biopsy Controller

    If the MAMMOMAT power is switched on, there is mains voltage present at the biopsy controller mains cord connector. Fig. 4 Biopsy controller Opdimar Register 5 SPB7-230.061.11 Page 4 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 51: Removal Of Biopsy Controller Cover

    If Led V2 (NVM error) and error message “Biopsy unit not NOTICE responding” appear after replacement of D200, follow the proce- dure described in Initialization of BC nvram on the D200 board on Page 9 - 2. Siemens AG Register 5 SPB7-230.061.11 Page 5 of 14 Opdimar Medical Solutions Rev.
  • Page 52 To replace the EPROMs of the board D200 in the biopsy unit, proceed as follows: Use PROM extractor to remove the EPROM. Install the new EPROM. Opdimar Register 5 SPB7-230.061.11 Page 6 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 53: Cables

    Z801 Printer M3000 printer/ M3000 stand (option) Service PC X885 Biopsy unit M3000 generator (M3000 stand) Fig. 7 Connection of cables Siemens AG Register 5 SPB7-230.061.11 Page 7 of 14 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 54: Biopsy Unit

    Smart card reader (not used in Opdima) 3.5-inch diskette drive (not used in Opdima) 5.25-inch CD drive Fig. 8 Front panel overview Opdimar Register 5 SPB7-230.061.11 Page 8 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 55 Audio module headphones connector, not used Audio module line-out connector, not used Audio module line-in connector, not used Audio module microphone connector, not used Fig. 9 Back panel overview Siemens AG Register 5 SPB7-230.061.11 Page 9 of 14 Opdimar Medical Solutions Rev.
  • Page 56: Ccd Camera

    Check of Opdimar AEC, see Check of Opdima AEC on Page 10 - 2. Check of resolution, see Check of resolution on Page 10 - 4. Opdimar Register 5 SPB7-230.061.11 Page 10 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 57: Reinstallation Of Software

    When a dialog box “Hostname” appears. Enter the original hostname (opdxxxx) and select continue. When a dialog box “Confirmation information” appears, select continue. The installation script is running. This will take approximately 20 minutes. The system reboots. Siemens AG Register 5 SPB7-230.061.11 Page 11 of 14 Opdimar Medical Solutions Rev.
  • Page 58: Installation Of Opdima Software Installation Cd-Rom (Mat. No. 66 33 700)

    Software upgrade and Restore hostname. NOTICE Restore hostname from most recently used backup MO disk. Opdimar Register 5 SPB7-230.061.11 Page 12 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 59 Select Restore and press Enter to copy the data to the hard disk. Set-up the printer once again after software reinstallation. Printer NOTICE information can not be restored. Control and if necessary activate or deactivate the screen lock function. Siemens AG Register 5 SPB7-230.061.11 Page 13 of 14 Opdimar Medical Solutions Rev.
  • Page 60: Final Procedures

    (x, y, z), calibrate the biopsy unit according to Calibration of the biopsy unit on Page 5 - 3. Turn screen lock on/off, according to Software upgrade, Miscellaneous on Page 5 - 18. Opdimar Register 5 SPB7-230.061.11 Page 14 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 61: Workstation

    Error transmitting targets to the biopsy The system has detected an error while trans- unit. mitting target to the biopsy unit. Check connec- tions. Try again. Siemens AG Register 5 SPB7-230.061.11 Page 1 of 4 Opdimar Medical Solutions Rev.
  • Page 62 If backup exists, restore backup data, see Supplement to the Restoring backup data in Instructions for Use MAMMOMAT 3000 - Opdimar Opdimar Register 5 SPB7-230.061.11 Page 2 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 63: Control Panel

    To resume Opdima mode; in the control dialog of the Opdima system, select Exit and choose the required mode (Stereo or Spot) in the Mode selec- tion dialog. Siemens AG Register 5 SPB7-230.061.11 Page 3 of 4 Opdimar Medical Solutions Rev.
  • Page 64 7 - 4 Messages This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 4 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 65: Fault Isolation Chart

    13. Exchange faulty unit. error log OK? 14. Problem still Contact your service present? support center System OK Fig. 1 Fault isolation chart Siemens AG Register 5 SPB7-230.061.11 Page 1 of 2 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24...
  • Page 66: Explanations

    Perform units test from the workstation service mode. Check the log for abnormal occurrences. Change faulty unit according to instructions. If the problem is still present, contact your service support center. Opdimar Register 5 SPB7-230.061.11 Page 2 of 2 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 67: General

    Log out and log in as “poweroff” to shut down the system. If necessary, power off the MO unit and turn it back on to be able to eject the disk. Then turn on the workstation again. Siemens AG Register 5 SPB7-230.061.11...
  • Page 68: Biopsy Unit

    Proceed as below to perform a ”dummy calibration”, that will update the checksum. The system MUST NOT be used with this ”dummy calibration”! NOTICE Calibrate with ”Biopsy Calibration” directly when the system is Opdimar Register 5 SPB7-230.061.11 Page 2 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 69 Press Ctrl + Alt + ! simultaneously. Confirm by pressing ok. It will now switch to custom behavior (the screen flashes for a second as the window manager is restarted). Perform a Calibration of the biopsy unit on Page 5 - 3. Siemens AG Register 5 SPB7-230.061.11 Page 3 of 14...
  • Page 70: Camera

    Power off and reboot Opdimar workstation. • Check cables. • Check that board D100/D101 is properly connected. • Change D100/D101. • Change camera. Opdimar Register 5 SPB7-230.061.11 Page 4 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 71: Workstation

    If no image on the monitor, check that input select is set to 1. NOTICE If removing the cover of the workstation main unit, check that no copper grounding clip comes loose. Siemens AG Register 5 SPB7-230.061.11 Page 5 of 14...
  • Page 72: Problems With The Centering Or Size Of The Displayed Image On The Monitor

    Follow the Instructions on the screen. Change back to the original language, if the customer was running another language than english. Opdimar Register 5 SPB7-230.061.11 Page 6 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 73: Tests

    Files included on the CD-ROM; Sun Blade 150 Getting Started Guide, Sun Blade 150 Service Manual with Sun Blade 150 ShowMe How Animations and Setting Up the Sun Blade 150 System Siemens AG Register 5 SPB7-230.061.11 Page 7 of 14...
  • Page 74: Image Quality

    Calibrate the CCD camera, move acrylic plastic plates after each calibration exposure. • If it is not disappearing, please contact the headquarter support center, HSC 24/USC, for investigation. Opdimar Register 5 SPB7-230.061.11 Page 8 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 75: Mammomat

    • Power off and reboot the Opdimar workstation. • Check biopsy controller, fuses, and power. • Check cables. • Check D707, must be installed. Siemens AG Register 5 SPB7-230.061.11 Page 9 of 14 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24...
  • Page 76: Network Problem

    • The license key to enable the DICOM (option) is written on the invoice and the dispatch note (if not, please contact Siemens-Elema AB mammography department in Solna, Sweden for further information). • To be able to receive the optional DICOM license key the mammography logistics department will request for the serial No.
  • Page 77: Printer

    • IP address from network administrator at the hospital. • When defining the printer in advanced service use PS for all postscript printers and XWD for Codonics printer. Siemens AG Register 5 SPB7-230.061.11 Page 11 of 14 Opdimar Medical Solutions Rev.
  • Page 78: Mo Unit

    Test of SCSI bus WARNING Test of SCSI bus can destroy the hard disk content. The software may have to be reinstalled. Opdimar Register 5 SPB7-230.061.11 Page 12 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 79 4. Check that the switch settings on the back of the MO is right configured. (see Fig. 3, Page 6 - 3) 5. Replace the SCSI-cable, MO unit and SCSI-board and enter the probe-scsi-all command between each replacements to isolate the fault to a specific unit. Siemens AG Register 5 SPB7-230.061.11 Page 13 of 14...
  • Page 80 9 - 14 Troubleshooting guide This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 14 of 14 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 81: Verifying The Calibration Of The Biopsy Unit

    Check that the biopsy unit displays do not show values more than ±1 mm. If they do, perform a biopsy calibration. After calibration of the biopsy unit, perform step 1. to 5. again. Fill in your values in the test protocol for repetitiveness (see Appendix 2). Siemens AG Register 5 SPB7-230.061.11 Page 1 of 6...
  • Page 82: Check Of Opdima Aec

    Opdimar. Because of this, no adjustments shall be done to these parameters before reading the AD center value, in order to get a correct reading. Opdimar Register 5 SPB7-230.061.11 Page 2 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 83: Performance Criteria

    Fig. 2 Pattern of “blocks” Siemens AG Register 5 SPB7-230.061.11 Page 3 of 6...
  • Page 84: Check Of Resolution

    Opdima) will be the limiting factor. For High Resolution mode the maximum theoretical resolution is > 20 lp/mm, and the x-ray tube (part of the MAMMOMAT) will be the limiting factor. Opdimar Register 5 SPB7-230.061.11 Page 4 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 85: Protective Earth Measurement

    Use a sharp-pointed tool to penetrate the finishing paint. Fill in your values in the test protocol for protective earth measurement (see Appendix 2). The value of the resistance measured must not exceed 0.1 Ω. Siemens AG Register 5 SPB7-230.061.11 Page 5 of 6...
  • Page 86: Ccd Camera

    Fill in your values in the test protocol for protective earth measurement (see Appendix 2). The value of the resistance measured must not exceed 0.1 Ω. Opdimar Register 5 SPB7-230.061.11 Page 6 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 87 Changes to previous version 11 - 1 Chap. Name of chapter Changes to previous version Completely revised. General The document is valid for ASW 3.1. Siemens AG Register 5 SPB7-230.061.11 Page 1 of 2 Opdimar Medical Solutions Rev. 02 07.03...
  • Page 88 11 - 2 Changes to previous version This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 2 of 2 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 89: Database Log File

    Stereo data: 0 Target side: SIN ----------------------------- Reading folder id: 877526913 patient id: 00-00002 Folder name: Folder type: spot examination Folder creation time: 877526913 Siemens AG Register 5 SPB7-230.061.11 Page 1 of 4 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24...
  • Page 90 Copied to MO disk #0 NOTICE! There are no images to save, see Page A2-1. Last modified 877524428 ----------------------------- Reading folder id: 877526913 Opdimar Register 5 SPB7-230.061.11 Page 2 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 91 Folder id 0: 877526913 ----------------------------- Reading patient id: 00-00003 Number of folders: 2 Folder id 0: 877527322 Folder id 1: 877567455 ----------------------------- Siemens AG Register 5 SPB7-230.061.11 Page 3 of 4 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 92 12 - 4 Appendix 1 This page intentionally left blank. Opdimar Register 5 SPB7-230.061.11 Page 4 of 4 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 93: Customer Specific Data

    Date Pattern Service Center Fictive patient ID Institution Name Department Name Test protocol repetitiveness Biopsy calculations ∆ ∆ ∆ Tester Object Siemens AG Register 5 SPB7-230.061.11 Page 1 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 94: Test Protocol Ccd Camera Calibration

    ________________ (27 kV grid, 26 kV non grid) Anode/filter material ________________ (Mo/Mo grid, Mo/Mo non grid) Mode ________________ (AEC mode grid, AEC mode non grid) Opdimar Register 5 SPB7-230.061.11 Page 2 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 95: Test Protocol Image Quality

    W/Rh Measured mean value, using ~ 80 mm calibration plexiglas Anode/Filter 26 kV 28 kV 30 kV 32 kV Mo/Rh W/Rh Siemens AG Register 5 SPB7-230.061.11 Page 3 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 96 W/Rh Measured mean value, using ~ 80 mm calibration plexiglas Anode/Filter 26 kV 28 kV 30 kV 32 kV Mo/Rh W/Rh Opdimar Register 5 SPB7-230.061.11 Page 4 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 97: Resolution

    Best resolution visible lp/mm Performance criteria: >=13 lp/mm Anode/filter combination: W/Rh Resolution: High Best resolution visible lp/mm Performance criteria: >=13 lp/mm Comments:_______________________________________ Siemens AG Register 5 SPB7-230.061.11 Page 5 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 98: Test Protocol Protective Earth Measurement

    Max. measured resistance Resistance limit ( Ω Biopsy controller cable duct Ω Measuring point Max. measured resistance Resistance limit ( Ω Opdimar Register 5 SPB7-230.061.11 Page 6 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 99: Customer Specific Data

    Date Pattern Service Center Fictive patient ID Institution Name Department Name Test protocol repetitiveness Biopsy calculations ∆ ∆ ∆ Tester Object Siemens AG Register 5 SPB7-230.061.11 Page 1 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 100: Test Protocol Ccd Camera Calibration

    ________________ (27 kV grid, 26 kV non grid) Anode/filter material ________________ (Mo/Mo grid, Mo/Mo non grid) Mode ________________ (AEC mode grid, AEC mode non grid) Opdimar Register 5 SPB7-230.061.11 Page 2 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 101: Test Protocol Image Quality

    W/Rh Measured mean value, using ~ 80 mm calibration plexiglas Anode/Filter 26 kV 28 kV 30 kV 32 kV Mo/Rh W/Rh Siemens AG Register 5 SPB7-230.061.11 Page 3 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 102 W/Rh Measured mean value, using ~ 80 mm calibration plexiglas Anode/Filter 26 kV 28 kV 30 kV 32 kV Mo/Rh W/Rh Opdimar Register 5 SPB7-230.061.11 Page 4 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...
  • Page 103: Resolution

    Best resolution visible lp/mm Performance criteria: >=13 lp/mm Anode/filter combination: W/Rh Resolution: High Best resolution visible lp/mm Performance criteria: >=13 lp/mm Comments:_______________________________________ Siemens AG Register 5 SPB7-230.061.11 Page 5 of 6 Opdimar Medical Solutions Rev. 02 07.03 CS PS 24 Service...
  • Page 104: Test Protocol Protective Earth Measurement

    Max. measured resistance Resistance limit ( Ω Biopsy controller cable duct Ω Measuring point Max. measured resistance Resistance limit ( Ω Opdimar Register 5 SPB7-230.061.11 Page 6 of 6 Siemens AG Service Rev. 02 07.03 CS PS 24 Medical Solutions...