Universal Needle Guide, Stainless
WARNING: Needle Guide Bracket kits are packaged non-sterile. Sterilize these
products prior to their first use. Refer to the sterilization procedures for the
Universal Needle Guide in the following pages.
Prior to sterilization, clean the universal needle guide bracket and insert(s)
using an enzymatic cleaner.
To clean:
1. Rinse the bracket and insert(s) with water.
2. Soak the bracket and insert(s) in an enzymatic cleaner.
Carefully follow the manufacturer's instructions.
3. Rinse the bracket and insert(s) with water to remove any debris and
remaining cleaner.
To sterilize:
Sterilize the bracket and needle guide insert(s) using a wrapped,
gravity-displacement steam sterilization at a temperature of 121ºC to
123ºC for an exposure time of 15 to 30 minutes.

Storage

Always clean and sterilize components used in a needle puncture or biopsy
procedure after each use.