Maintenance Dialog Box Stress Echo

Maintenance Dialog
Box Selection
Reference Image
On
Off
Acquire Mode
Manual ROI
Fixed ROI
Fullscreen
Image Compression
High
Medium
Overlay
Name of level
Name of view
Heartrate
Time stamp
acquisition
Timespan since
start of level
Compression Ratio
WMS Type
ASE 16 – 4 Scores
1
ASE 17—4 Scores
[ 1 ]
I N S T R U C T I O N S
Description
Enables display during acquisition of the reference image (the view in the initial
phase that corresponds to the selected view).
Note: The reference image is displayed if the initial stage was non-continuous.
Disables display during acquisition of the reference image (the view in the initial
phase that corresponds to the selected view).
Enables display of ROI (region of interest) after the Stress Echo protocol is
selected. The ROI can be resized and/or repositioned as required. The ROI
determines the portion of the image area to be saved.
Implicitly sets the ROI to exclude a portion of the bottom of the image; for each
image saved, the system captures all of the image area except the excluded
portion at the bottom of the image. No ROI is displayed after the Stress Echo
protocol is selected.
Implicitly sets the ROI to include the gray bar in addition to the image area; for
each image saved, the system captures the entire image area and the gray bar.
No ROI is displayed after the Stress Echo protocol is selected.
Sets image compression during acquisition to High (12:1).
Sets image compression during acquisition to Medium (8:1).
Displays the phase label on the loop in the Stress Echo screen.
Displays the view label on the loop in the Stress Echo screen.
Displays the heart rate at which the loop was acquired on the loop in the
Stress Echo screen.
Displays the time of acquisition on the loop in the Stress Echo screen.
Displays the time elapsed from the beginning of the phase to this loop's
acquisition on the loop in the Stress Echo screen. This value can be used in an
exercise-based stress echo protocol to indicate the time elapsed since the patient
began exercising.
Displays the compression ratio of the corresponding loops on the loop in the
Stress Echo screen.
Configures the system to display the indicated type of WMS graphics
(16-segment or 17-segment).
Configures the system to display the following WMS scores for assignment:
X – Cannot evaluate
1 – Normal
2 – Hypokinetic
3 – Akinetic
4 – Dyskinetic
F O R
U S E
3 S y s t e m C o n t r o l s
3 - 5 9