System Surfaces
The following instructions describe cleaning the surface of the ultrasound
system, including the trackball and transducer holder.
To clean the surface of the ultrasound system:
1. Power off ( ) the ultrasound system and unplug the power cord from
the power outlet.
2. Use a clean gauze pad or lint-free cloth, lightly moistened with a mild
detergent, to wipe the surface of the ultrasound system.
Take particular care to clean the areas near the trackball and the slide
controls. Ensure these areas are free of gel and any other visible
residue.
Ensure that cleaning solution does not seep into the control panel, LCD,
keyboard, or any other openings.
3. After cleaning, use a clean, lint-free cloth to dry the surface.
4. After cleaning, reconnect the ultrasound system power cord into the
power outlet.
To clean the holders for transducers and coupling gel:
1. Remove the holder from the ultrasound system:
Reach under the holder to locate the tab on the holder. The tab extends
below the point of attachment to the ultrasound system. Squeeze the
tab towards the holder and pull the holder downward.
2. Clean the holder under running water, using a mild detergent and dry
with a lint-free cloth.
3. Reattach the holder to the ultrasound system:
Align the support on the holder directly below the point of attachment
on the ultrasound system and firmly push upwards until the holder
snaps into place.