Transmit Power Display

The transmit power range is from –20 dB to 0 dB. Selecting 0 dB, or 100%, in
combination with other system controls or functions generates the maximum
acoustic intensity and mechanical index for each transducer, where:
2
I
:
720 mW/cm
SPTA.3
Use the system presets to determine whether the power value displays on
the image screen in decibels, dB, or as a percentage of the maximum
allowable power.
0 dB
-3 dB
-6 dB
100%
71%
50%
Comparison of decibels to percentages.
When the power level is changed, the system briefly highlights the power
value to indicate the change.
.... .... ....
+
+
C 5/2/ 26
Location of power display on the image screen.
[ 1 ]
I N S T R U C T I O N S
and MI
1.9
-9 dB
-12 dB
-15 dB
35%
25%
18%
....
.... .... .... ....
+
+
+
+
+
+
F O R
U S E
-18 dB
-21 dB
-24 dB
13%
9%
6%
2 S a f e t y a n d C a r e
System Reference
System Presets
Ch 3
F4
General
Beep Volume
Tx Power
Display Format
2 - 1 1