3 S y s t e m C o n t r o l s

ECG LCD Selections

When ECG is active, you can access the related LCD selections by rotating
the FUNCTION SELECT control on the LCD panel to highlight the ECG tab
on the LCD.
Use the system presets to enable the ECG function to automatically activate
when an exam type is selected.
LCD Selection
On/Off
CH 1 Delay
Position
Gain
CH 2 Delay
External
Single
Trigger
3 - 4 6
Description
Activates (On) the ECG function.
Selects the channel 1 trigger point in the cardiac cycle
where the system updates the image from ventricular
depolarization, or the 'R' value of the ECG tracing.
Adjusts the vertical position of the ECG trace on the
image screen.
Increases or decreases the trace amplitude.
Selects the channel 2 trigger point in the cardiac cycle
where the system updates the image from ventricular
depolarization, or the 'R' value of the ECG tracing.
Activates (On) the external ECG function.
Selects Single or Dual trigger within one cardiac cycle.
Enables ECG-triggered acquisition of 2D-mode data.
F4
Default Setting
ECG
Settings
On, Off
0 ms to 990 ms in
increments of 10
0 to 31 in
increments of 1
0 to 15 in
increments of 1
0 ms to 990 ms in
increments of 10
On, Off
Sngl, Dual
On, Off
[ 1 ]
I N S T R U C T I O N S
F O R
U S E