Setting COC inlet parameters

Advanced Operation Manual
Track oven mode
Press [Front Inlet] or [Back Inlet].
1
Press [Mode/Type] and select Track oven.
2
There is no setpoint for Track oven mode.
Ramped temperature mode
Press [Front Inlet] or [Back Inlet].
1
Press [Mode/Type] and select Ramped temp.
2
Enter a value for Temp. This is the starting temperature.
3
Enter an Init time. This is the length of time the inlet will
4
stay at the starting temperature after a run has begun.
Enter Rate 1. This is the rate at which the inlet will be
5
heated or cooled. A Rate of 0 halts further programming.
Enter Final temp 1. This is the inlet temperature at the end
6
of the first ramp.
Enter Final time 1. This is the number of minutes the inlet
7
holds Final temp 1.
To enter a second (or third) ramp, scroll to the
8
appropriate Rate line and repeat steps
3
Inlets
5
through 7.
95