4

Installation

4.1

Electrical connection

4.1.1

Electrical connections between the transmitter and the flowmeter sensor

Connecting transmitter ME2 to flowmeter sensor MS2
92 91 85 86 87 88 89 90 93 94 95 96
11
1
2
3
4
Fig. 2
91 / 92
Driver
93 / 94 / 95 / 96
Temperature
85 / 86
Sensor 1
87 / 88
Sensor 2
CI/FCM2000-EN
5
6
7 8 9 10
CoriolisMaster FCM2000
11
11
11
11
1 Red
2 Brown
3 Green
4 Blue
5 Gray
6 Violet
7 White
8 Black
9 Orange
10 Yellow
11 "PA" equipotential bonding. When connecting transmitter to
flowmeter sensor MS26, transmitter also has to be connected to
"PA".
Installation
M 22, 23, 24, 25
E
ME27
Me28
MS21
Ms26
G00389