CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
Lay seal wire
4
Fig. 17: Example of FCX100 and FCX400 sensor in remote mount
design
3.
Continue to twist the seal wire
hole screw
2
. Guide one end of the wire through the
hole in the screw, guide the other end near the screw.
Tightly twist the seal wire
screw
2
to secure the screw.
4.
Continue to twist the seal wire
hole screw
3
. Secure the front head hole screw
described previously.
5.
Tightly twist both ends of the seal wire after the front
head hole screw
3
over a length of approx. 5 cm.
6.
Place the seal
4
approx. 5 cm from the front head hole
screw
3
onto the ends of the seal wire and press
together using sealing pliers. Cut off any excess
wire.
G12365
up to the rear head
1
1
right behind the head hole
up to the front head
1
3
1
3.5

FCT400 sensor with dual-compartment housing

NOTICE
Before the device is sealed, any settings must be
completed and the write protection switch must be set to
"write protection active" (see chapter 'Activating the
hardware write protection' on page 3 ).
1
Scope of delivery
2
3
1
2
3
Fig. 18
Item
Designation
1
as
Holding rod
2
Locking cylinder
3
Head hole screw
Seal
Seal wire
seal
EN – 11
G12350
Quantity
2
2
2
1