Data
Conversion
Data
sent
from
the
HP
54501A
is
raw data and must be scaled for useful
interpretation. The
values used to interpret the data are the X and Y
references,
X
and
Y
origins, and
X
and
Y
increments. These
values are
read from the waveform preamble.
Conversion from
The formula
to convert
a
data value from waveform memories 1-4 to
a
Data
Value
tO
voltage
value
is:
Voltage
voltage
=
[(data
value
-
yreference)*yincrement]
+
yorigin
Conversion from
The
time value
of
a
data point can be determined
by
the position of the
Data Value to Time
data point. As an example, the
third data point
sent
with
XORIGIN
=
16
ns,
XREFERENCE
=
0,
and
XINCREMENT
=
2ns. Using
the
formula:
time
=
[(data point number
•
xreference)*xincrement]
+
xorigin
would
result in the
following
calculation:
time
=
((3
-
0)
*
2
ns]
+
16
ns
=
22
ns.
Waveform Subsystem