VPP
VPP
command/query
The
MEASURE:
VPP
command places the instrument in the continuous
measurement mode and
starts
a
VPP
measurement.
The VPP
query measures the maximum and minimum voltages for the
selected source, then calculates the peak-to-peak voltage and outputs that
value.
The
peak-to-peak voltage
(Vpp)
is
calculated with the formula:
Vpp
=
Vmax
-
Vmin
where Vmax and Vmin
are the
maximum
and minimum voltages present
on the selected
source.
Command
Syntax:
:MEASure:vpp
Example:
OUTPUT 707;":MEAS:vpp"
Query Syntax:
:MEASure:vpp?
Returned Format:
[:MEASure:vpp]
<vaiue>
<NL>
Where:
<
value
>
::
=
voltage peak to peak (exponential
-
NR3 format)
Example:
DIM Vp$[50]
OUTPUT 707;":MEAS:VPP?"
ENTER 707;Vp$
PRINT Vp$