SOURce
SOURce
command/query
Command Syntax:
The
:MEASURE:SOURCE
command
selects
the source(s)
for the
measurements.
The
source
specified will become
the source
for the
Measure
subsystem commands.
Two
sources can be
specified
with
this
command.
All
measurements
except
DELAY
are made on the first specified
source.
The
DELAY
measurement will
use
two
sources
if
two
have
been specified.
If
only one
source
is
specified
the
DELAY
measurement will
use that source
for both
of
its
parameters.
The
SOURCE
query returns
the current source selection.
If
the specified
sources are
different both will be returned, otherwise
one source
will be
returned.
:MEASure:SOURce <source1
>[,<source2>]
Where:
Example:
Query Syntax:
Returned Format:
<source1
>
and
<source2> ::=
{CHANnel{1
[
2
|
3
| 4} |
FUNCtion{1
|
2}
WMEMory{1
|
2
|
3
| 4}}
OUTPUT
707;":MEASURE:SOURCE CHANNEL1, WMEMORY1"
:MEASure:SOURce?
[:MEASure:SOURce] <source1
>[,<source2>]<NL>
Where:
<souree1
>
and
<source2> ::=
{CHANnel{1
|
2
|
3
| 4} |
FUNCtion{1
| 2}
WMEMory{1
|
2
|
3
| 4}}
Example:
DIM Src$[50]
OUTPUT
707;":MEAS:SOUR?"
ENTER 707;Src$
PRINT Src$
HP 54501A
Programming
Measure Subsystem
14-39