Complete - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

COMPIete
COMPIete
command/query
The
:ACQUIRE:COMPLETE
command
specifies the completion criteria
for
an acquisition.
The parameter determines
what percentage of the time
buckets need to be
"full"
before an acquisition
is
considered complete.
If
you are in the
NORMAL
mode the instrument only needs one data bit per
time bucket for that time bucket to be considered full.
In
order for the
time bucket to be considered full
in
the
AVERAGE
or
ENVELOPE
modes
a
specified number of
data points
(COUNT)
must be acquired.
The
range for the
COMPLETE
command
is 0
to 100
and indicates the
percentage
of time
buckets
that must be
"full"
before the acquisition
is
considered complete.
If
the complete value
is
set to
100%,
all
time
buckets
must contain data for the acquisition to be considered complete.
If
the
complete
value
is set
to 0 then one acquisition cycle
will take
place.
The
COMPLETE
query returns the completion criteria
for the currently
selected mode.
Command
Syntax:
:ACQuire:COMPIete
<comp>
Where:
Example:
Query Syntax:
Returned Format:
Example:
<comp> ::=
Oto 100 percent
OUTPUT
707;":ACQUIRE:COMPLETE
85"
:ACOuire:COMPIete?
[:ACOuire:COMPIete]
<compxNL>
Where:
<comp> ::=
Oto 100 (integer-
NR1
format)
DIM
Cmp$[50]
OUTPUT
707;":ACQUIRE:COMPLETE?"
ENTER
707;Cmp$
PRINT
Cmp$
Acquire Subsystem