CONDition
CONDition
command/query
The
:TRIGGER:CONDITION
command
is
valid in the
PATTERN,
STATE, DELAY,
and
TV
trigger
modes.
The function of
the
CONDITION
command in each
of
these modes
is
described below.
Time
values
entered using
this
command are rounded to the nearest
10
ns.
In
the Pattern
Trigger
Mode
the valid arguments
for
the
CONDITION
command are
ENTER,
EXIT,
GT, LT, RANGE.
In
the
Pattern Trigger Mode the
CONDITION
command
is
used to
specify
if
the trigger will be generated upon entry to the specified
logic
pattern, upon exiting the specified logic pattern, or if the pattern must be
present for
a
specified amount of time.
The time
in the
pattern trigger
mode
can be
specified to be greater than
a
value
(GT),
less
then
a
value
(LT),
or between two values
(RANGE).
These are the same settings that are specified using
the
front panel
"when"
key
in the
Pattern Trigger Mode.
In
the State
Trigger Mode
the valid parameters for
the
CONDITION
command are
TRUE
(is
present) and
FALSE
(is
not
present).
In
the Delay
Trigger Mode the
CONDITION
command
is
valid when
PATTERN
or
STATE
is
selected
as
the qualifier.
All
arguments for this
command that are valid in the Pattern or
State
Trigger Modes are valid
here.
In
the
TV
Trigger Mode
the
CONDITION
command
is
used to
set
the
range
of time
values for the trigger to occur.
This command
is
only valid
in the "user defined"
mode.
The
CONDITION
query returns the currently
selected condition, for the
currently selected mode.