RANGe
RANGe
command/query
The
:TIMEBASE:RANGE
command
sets
the full
scale
horizontal time
in
seconds.
The
RANGE
value
is
ten times the time per
division.
The
RANGE
query returns the current range
value.
Command
Syntax:
:TIMebase:RANGe
<
range
>
<
range
>
::= 20ns
to
50s
in
a
1,2,5 sequence"
Example:
OUTPUT
707;":TIM:RANG 100 MS"
Query Syntax:
:TIMebase:RANGe?
Returned Format:
[:TIMebase:RANGe]
<
range
>
<
NL>
Where:
<
range
>
::
=
20
ns to
50
s
(exponential
-
NR3 format)
Example:
DIM Rng$[50]
OUTPUT 707; ":TIMEBASE:RANGE?"
ENTER 707;Rng$
PRINT
Rng$
HP 54501