HP 54501A Programming Reference Manual page 217

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

COMPare
Query Syntax:
:MEASure:COMPare?
<
measurement
>
Where:
<
measurement
>
:;=
{RISetime
|
FALLtime
|
FREQuency
|
PERiod
|
PWIDth
|
NWIDth
|
VAMPIitude
|
VBASe
|
VTOP
|
VPP
|
VAVerage
|
VMAX
|
VMIN
|
VRMS
|
DUTycycle
|
DELAY}
Returned Format:
[:MEASure:COMPare]
<
measurement
>, <upperjmit>,
<lower_limit>
<NL>
Example:
DIM Cmp$[50]
OUTPUT
707;"MEASure:COMPare?
VPP"
ENTER 707;
Cmp$
PRINT
Cmp$
For example,
the sequence required to do
a
limit
test
on frequency
is:
OUTPUT
707;":MEASURE:FREO"
!Select
measurement
OUTPUT
707;":MEASURE:COMPARE FREO,1000HZ,10HZ"
'Set measurement limits
OUTPUT 707;":MEASURE:LlMITTEST
MEASURE"
"Start
test
Note
The only way to
see
if
a
limit
test
failure
has
occurred over
the
bus
is
by checking
if
bit 3
of
the status
byte
is set
to
a
1.
HP
54501A
Programming
Measure Subsystem
14-13