Cursor - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

query
The
:MEASURE:CURSOR
query returns the time and voltage
values
of
the specified marker
as
an
ordered pair
of
time/voltage
values.
When
the
CURSOR
query
is sent,
no measurement
is
made
and the
cursors
are not moved.
If DELTA
is
specified:
the instrument returns the value
of
delta
V
and delta
T.
If
START
is
specified:
the positions
of
the
start
marker and
VMarker
1
are returned.
If
STOP
is
specified:
the positions
of
the stop
marker
and
VMarker
2
are returned.
Query Syntax:
:MEASure:CURsor? {DELTa
I
STARt
|
STOP}
Returned Format:
[:MEASure:CURSoc]
<
time
>,<
voltage
><NL>
Where:
<time
>
::
=
delta
time,
start
time
or stop
time
<
voltage
>
::
=
delta voltage, VMarker
1
voltage
or
VMarker
2
voltage
Example:
OUTPUT 707;":MEAS:SOURCE
CHAN1"
OUTPUT 707;":MEAS:CURSOR?
START"
ENTER 707;Tme,Vlt
PRINT Tme.VIt
Measure Subsystem
14-14
HP
54501A
Programming

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents