TER
TER
(Trigger Event Register)
query
The
TER
query
allows
the Trigger Event Register to be read.
When
the
Trigger Event Register
is
read
it is
cleared.
A
one indicates
a
trigger
has
occurred.
A
zero indicates
a
trigger
has
not occurred.
If
a
trigger
event is not
found and the sweep
is
auto-triggering
this
bit will
not be
set.
A
Service
Request
(SRQ)
can
only be generated when the bit transitions
from
0
to
1,
therefore the
bit must be cleared each
time you would like
a
new Service Request to be generated.
Query Syntax:
:TER?
Returned Format:
[:TER]
{i |
0}
<
NL
>
Example:
OUTPUT
707;":TER?"
ENTER
707;Trg^event$
PRINT
Trg_event$