Averaging
Mode
The
:ACQUIRE:TYPE AVERAGE
command
sets
the
HP
54501A
to the
Averaging mode.
COUNT
can be
set
in
AVERAGE
mode
by
sending the
:ACQUIRE:COUNT
command followed
by
the number
of
averages.
In
this
mode the
value
is
rounded to
the nearest
power of
2.
It
determines
the number of averages that must be acquired.
To
activate the averaging
mode from the front
panel, select the
Display
menu, then
select
Average. Changing the number of
averages
changes the
COUNT
value.
Envelope Mode
The
:ACQUIRE:TYPE ENVELOPE
command
sets
the
HP
54501A to
the Envelope mode.
A
count value can be
set
in the envelope mode. This value determines the
number of
values to be used, at each
time point, when constructing
the
envelope.
The
count value cannot be set
from the front panel.
To
activate the
Envelope mode from
the
front panel,
select the
Display
menu,
then select the Envelope mode.
Acquire Subsystem
8-2
HP 54501A
Programming