HOLDoff
HOLDoff
command/query
The
:TRIGGER:HOLDOFF
command
is
valid in the Edge, Pattern,
State,
and
TV
Trigger
Modes. This command will allow
a
holdoff by time
value to be entered.
The
HOLDOFF
query
returns the value
of
the
holdoff
for the current
mode.
Command
Syntax:
:TRIGger:HOLDof( TIME,
<
holdoff_value
>
Where:
Examples:
Query Syntax:
Returned Format:
Example:
<
holdoff_value
>
::
=
40
ns to 320 ms rounded to nearest
20
ns increment.
OUTPUT 707;":TRIGGER:HOLDOFF TIME.216
US"
OUTPUT 707;":TRIGGER:HOLDOFF TIME.2.16E-4"
:TRIGger:
HOLDoff?
[:TRIGger:HOLDoff] <holdoff_value>
<NL>
Where:
<
holdoff_value
>
::
=
40
ns to
320 ms (exponential
-
NR3 format)
DIM Ho$[50]
OUTPUT
707;":TRIGGER:HOLD?"
ENTER 707; Ho$