SETup
SETup
command/query
The
:SYSTEM:SETUP
command
sets
the
HP
54501A
as
defined
by
the
data in the learn string
sent
from the controller. The
setup
string
contains
1024 bytes
of
setup data.
The 1024 bytes
does not include the header or
"#800001024".
The
SETUP
query outputs
the current
HP
54501A
setup in the
form of
a
learn string to the controller. The
SETUP
query operates the same
as
the
*LRN?
query.
The
learn string
is sent
and received
as
a
binary
block
of
data.
The
format
for the data
transmission
is
the
#
format defined
in
the
IEEE
488.2
specification.
Command
Syntax:
:SYSTem:SETup
<setup>
Example:
OUTPUT 707;":SYSTEM:SETUP
<
setup
>"
Where:
<setup> ::= ^800001024
<
setup
data
string
>