Range - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

RANGe
RANGe
command/query
The
:CHANNEL
<
N
>
:RANGE
command
defines the full scale vertical
axis
of
the
selected channel.
The
RANGE
for channels
1
and
4 can be set
to
any value from 40
mV
to 40
V,
when using
1:1
probe attenuation.
The
RANGE
for
channels 2 and 3 can
be
set
to
any value
from 800
mV
to
4.0
V,
when using
1:1
probe attenuation.
If
the probe attenuation
is
changed, the range value
is
multiplied
by
the probe attenuation factor.
The
RANGE
query returns the
current
range setting for the
specified
channel.
Command
Syntax:
:CHANnel<N>:RANGe
<
range
>
Where:
<N>
::=
1,2,3,or4
<
range
>
::=
range value
Examples:
:OUTPUT 707;"CHANNEL1
:RANGE
.60"
:OUTPUT
707;"CHANNEL2:RANGE
1.2
V"
:CHANnel<N>:
RANGe?
[:CHANnel<N>:RANGe]
<rangexNL>
Where:
Query Syntax;
Returned Format:
<N>
::=
1,2,3,or4
<
range
>
::
=
range value (exponential
-
NR3 format)
Example:
DIM Rng$[50]
OUTPUT
707;":CHAN2: RANGE?"
ENTER 707;
Rng$
PRINT Rng$
Channel Subsystem