Level - HP 54501A Programming Reference Manual

Digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

LEVel
LEVel
command/query
The
:TRIGGER:LEVEL
command
sets
the trigger level voltage
of
the
active trigger.
This command can be
sent
in any mode, however only two
separate levels are stored.
One
value
is
kept for the
TV
Trigger Mode
and another value
is
kept for
all other modes.
If
you are in the Pattern
Trigger Mode and
set
a
trigger
level value, that level will also be used
for
the Edge, State, and
Delay Trigger Modes.
The
LEVEL
query
returns the trigger
level
of
the
current trigger mode.
Command
Syntax:
:TRIGger:LEVel
<
level
>
Where:
<
level
>
::= -140.0
volts to
+140.0
volts
(maximum value depends on volts/division
and
offset settings).
Examples:
OUTPUT 707;":TRIGGER:LEVEL
.30"
OUTPUT 707;":TRIGGER:LEV
300MV
OUTPUT 707;":TRIG:LEV
3E-1"
Refer
to chapter 3
for the
syntax
of
using values with multipliers.
Query
Syntax:
:TRIGger:LEVel?
Returned Format:
[:TRIGger:LEVel]
<
level
>
<
NL>
Where:
<
level
>
::
=
trigger level
in
volts (exponential
NR3
format)
Example:
DIM Tlevel$[30]
OUTPUT
707;":TRIGGER: LEVEL?"
ENTER
707;Tlevel$
PRINT
Tlevel$