OVERshoot
OVERshoot
command/query
The
:MEASURE:OVERSHOOT
command places the instrument in the
continuous measurement mode and
selects
the Overshoot measurement.
The
OVERSHOOT
query measures and outputs the overshoot
of
a
selected signal. Overshoot measures the first edge on screen with the
following algorithm:
if
the first edge on screen
is
rising
then
overshoot
=
(Vmax
-
Vtop)/VampIitude
else
overshoot
=
(Vbase
Vmin)/Vamplitude
Command
Syntax:
:MEASure:ovERshoot
Example:
OUTPUT 707;";MEAS:OVER"
Query
Syntax:
:MEASure:ovERshoot?
Returned Format:
[:MEASure:ovERshoot]
<value>
<NL>
Where:
<
value
>
::
=
ratio
of overshoot to Vamplitude (exponential
-
NR3 format)
Example:
DIM Ow$[50]
OUTPUT 707;":MEASURE:OVERSHOOT?"
ENTER 707;0vr$
PRINT Ovr$
Measure Subsystem