LINE
LINE
command/query
The
:TRIGGER:LINE
command
is
valid in the
TV
Trigger Mode when
the
STANDARD
selected
is
525 or 625.
If
one
of
the these standards
is
selected when the
TV
Trigger Mode
is
entered the line value will be
set
in
that standard and selected field.
The
LINE
query returns the current line
of the
selected standard.
Command
Syntax:
:TRIGger:LINE
<line_number>
Where:
<line_number>
::
=
1
to
625 (depends
on STANDARD
and FIELD selection).
Example:
OUTPUT 707;":TRlG:LlNE
22"
Query Syntax:
:TRIGger:LINE?
Returned Format:
[:TRIGger:LINE]
<line_numberxNL>
Where:
<
linejiumber
>
::
=
1
to
625
(depends on STANDARD
and FIELD
selection).
(integer
-
NR1
format)
Example:
DIM Ln$[50]
OUTPUT 707;":TRIGGER:LINE?"
ENTER 707; Ln$
PRINT
Ln$