VERSUS
VERSUS
command
The
:FUNCTION
<
N
>:
VERSUS
command
allows X
vs
Y
displays with
two operands. The
first
operand defines
the
Y axis
and the second defines
the
X axis.
The
Y axis
range and offset
is
initially equal to that
of
the first
operand and
can be
adjusted with
the
FUNCTION
<
N
>
:RANGE
and
FUNCTION
<
N
>
:OFFSET
commands.
The
X axis
range and offset
is
always equal to that
of
the second
operand.
It
can only be changed
by
changing the vertical settings of
the second
operand. This will
also
change the
Y axis
vertical sensitivity and offset.
Command
Syntax:
:FUNCtion<N>:
VERSUS
<Y_operand>,<X_operand>
Where:
<N>
::=
1
or
2
<Y_operand> ::=
{CHANneh
|
CHANnel2
|
CHANnel3
|
CHANnel4
[
WMEMory-l
|
WMEMory2
|
WMEMoryS
| WMEMory4}
<X_operand> ::=
{CHANnel-l
|
CHANnel2
|
CHANnel3
|
CHANnel4
|
WMEMory-l
|
WMEMory2
|
WMEMory3
|
WMEMory4}
Example:
OUTPUT 707;":PUNCTION2:VERSUS CHAN1,CHAN2"
HP 54501A
Programming
Function Subsystem
12-11