CDMA Channel Level Calibration Procedure

Approximate time: 8 minutes
1. Load the PCB_CAL procedure.
1. Press the TESTS key.
2. Select ROM from the list of choices for the
Select Procedure Location field.
3. Select PCB_CAL from the list of choices for
the Select Procedure Filename field.
2. Run the PCB_CAL Procedure.
1. Select the Run Test field.
2. Disconnect the mobile station from the Test
Set's (or PCS Interface's) RF IN/OUT port.
(The Test Set beeps several times and displays
the message to disconnect all front-panel
cables. You only need to disconnect the RF
input from the mobile.)
3. When the PCB_CAL procedure has completed,
cycle power to the Test Set.
At one point in the procedure, the Test Set will beep and the message "Direct latch write occurred. Cycle power
when done servicing" will appear. This is normal. Cycle power when the procedure has completed.
Chapter 1, Calibrate the Test Set
CDMA Channel Level Calibration Procedure
TESTS (Main Menu)
Select Procedure Location:
ROM
Select Procedure Filename:
PCB_CAL
TESTS (Main Menu)
Select Procedure Location:
ROM
Select Procedure Filename:
PCB_CAL
RUN TEST
21