Some Detector and Video Bandwidth settings will raise the Auto Sweep
Acquisition value greater than 1. In these cases, manually setting Sweep
Acquisition lower than the Auto value may have NO effect.
NOTE
Measurement speed specifications do NOT apply in Temperature Control Mode.
Learn more on page 13.

Zero Span Measurements

When setting the frequency span to Zero, there is NO spectrum of frequencies to
display, so the X-axis units becomes Time. The SA becomes like a tunable
oscilloscope, with the center frequency being the frequency of interest. This
capability is useful for analyzing modulation characteristics, such as pulsed
measurements.
GSM signal, framed data format, timeslot 0 and 3 on. Sweep Time is set to
approximately the frame interval. Press Single several times until the
waveform section of interest is viewable and stable. Then markers can be
used to measure the timeslot width and interval as shown.
How to set Zero Span
 Set
How to change Sweep Time in Zero Span.
 Press Sweep 3
 Then
SA (Spectrum Analyzer) Mode
o
# is shown in front of the actual sweep time to indicate a manual setting.
Center
to the frequency of interest
then set
Span
to 0 Hz
o
or select
Zero/Last
Span. Each key press toggles the frequency span
o
between 0 Hz and the last specified frequency span.
SweepTime