Zero Span Measurements - Agilent Technologies FieldFox N9913A User Manual

Hide thumbs Also See for FieldFox N9913A:
Table of Contents

Advertisement

Some Detector and Video Bandwidth settings will raise the Auto Sweep
Acquisition value greater than 1. In these cases, manually setting Sweep
Acquisition lower than the Auto value may have NO effect.
NOTE
Measurement speed specifications do NOT apply in Temperature Control Mode.
Learn more on page 13.

Zero Span Measurements

When setting the frequency span to Zero, there is NO spectrum of frequencies to
display, so the X-axis units becomes Time. The SA becomes like a tunable
oscilloscope, with the center frequency being the frequency of interest. This
capability is useful for analyzing modulation characteristics, such as pulsed
measurements.
GSM signal, framed data format, timeslot 0 and 3 on. Sweep Time is set to
approximately the frame interval. Press Single several times until the
waveform section of interest is viewable and stable. Then markers can be
used to measure the timeslot width and interval as shown.
How to set Zero Span
 Set
How to change Sweep Time in Zero Span.
 Press Sweep 3
 Then
 Enter a value using the numeric keypad.
SA (Spectrum Analyzer) Mode
o
# is shown in front of the actual sweep time to indicate a manual setting.
Center
to the frequency of interest
Span
then set
to 0 Hz
o
Zero/Last
Span. Each key press toggles the frequency span
or select
o
between 0 Hz and the last specified frequency span.
SweepTime
93

Advertisement

Table of Contents
loading

Table of Contents