2-Port Transmission Measurement - Agilent Technologies FieldFox N9912A User Manual

Rf analyzer
Hide thumbs Also See for FieldFox N9912A:
Table of Contents

Advertisement

1. Attach the shortest cable to the FieldFox RF OUT connector. This cable is the
2. Press Meas Setup 4 then press Zero.
3. Remove the reference cable from the FieldFox RF OUT connector.
4. Connect an unterminated, longer cable to the FieldFox RF OUT connector.
5. Observe the displayed phase difference between the reference cable and the
6. Carefully trim the cable until the phase shift reads zero. The attached
7. Repeat steps 4 through 6 for the remaining cables to be trimmed.
2-Port Transmission Measurements
A 2-Port Transmission measurement is used for measuring electrical length,
insertion loss, gain, or isolation of a device. The FieldFox signal source is
transmitted out the RF OUT connector, through the DUT, and into the RF IN
connector. Both ends of the DUT must be connected to the FieldFox.
NOTE
Only Magnitude, NOT Phase, is displayed.
In the following image, the gain of an amplifier is being measured.
2-Port Transmission measurement of an amplifier
How to make a 2-Port Transmission Measurement
1. Press Mode then
2. Press Freq/Dist and enter the measurement frequency.
3. Connect the RF Out and RF In test port cables to form a Thru connection as
4. Press Meas Setup 4 then Zero.
5. Connect the DUT as in the diagram above.
6. The displayed magnitude value is the gain of the amplifier. To measure
VVM (Vector Voltmeter) Mode
reference cable. Leave the end of the reference cable open (unterminated).
attached cable.
cable's electrical length is now matched to the reference cable.
VVM
in the image below. If an adapter is needed to make the connection, the
measurement results will be in error by the amount of loss in that adapter.
isolation of the amplifier, reverse the connection to the amplifier (RF Out to
the amplifier output).
then
2-Port
Transmission.
119