How to perform a single sweep while in Hold
 Press Sweep 3.
 Then choose one of the following:

Preset

The Preset key restores the FieldFox measurement settings to a known condition.
You can preset the entire instrument, or just the current Mode (CAT, SA, Power
Meter, NA.)
When Preset, the FieldFox performs a calibration for use in CAT and NA Modes.
Learn more on page 51.
The following settings always survive Preset and Mode Preset:
 External Reference state (page 150)
 SA Units (page 63)
 SA Correction states (page 66)
 VVM Display Resolution (page 116)
 Marker Peak Excursion and Threshold (page 128)
 Volume Control
How to Preset the FieldFox
 Press Preset .
 Then choose one of the following:
Volume Control
System volume control is a FieldFox global setting which sets speaker volume for
ALL success beeps, key press clicks, and individual features that use the
FieldFox speaker, such as SA Mode Tune and Listen.
The volume control setting remains through a FieldFox Preset. To cause your
volume control setting to remain through a FieldFox shutdown, save the setting
as a Preference. Learn how on page 146.
How to set FieldFox Volume Control
 Press System 7
System Settings
Automatically sets Continuous OFF (Hold) and causes FieldFox to
Single
o
make ONE measurement sweep, then hold for the next Single key press.
SA Mode ONLY – With
o
also causes a single sweep.
Continuous ON OFF
o
when battery power is not critical.
Returns ALL Modes to the factory default settings.
Preset
o
Returns only the current Mode to the factory default settings.
Mode Preset
o
The current settings for all other Modes remain.
each successive Run/Hold press
Continuous OFF
Makes continuous sweeps. This is the typical setting
143