Battery Saver
When the battery saver is OFF, the source power is left ON at the end of the
sweep which consumes more power. This feature is used in CAT, NA and VVM
modes.
How to set Battery Saver
 Press System
 Then
 Then

System Configuration

Options (Licensing)
You can view and install options on your FieldFox. See all available options on
page 7.
How to view the Installed Options
 Press System
 Then
 Then
 Then
The currently installed options are listed.
How to Install Options
A .lic file must already be on a USB Flash Drive. To learn how to obtain a .lic file,
visit:
Insert the USB Flash Drive in the FieldFox.
 Press System.
System Settings
Save Current Settings
saved.
Reset Preferences
Preferences
Battery Saver
ON – Default setting. To conserve battery power, the RF source power is
o
turned OFF at the end of each CAT, NA, and VVM mode sweep. This
behavior does not impact FieldFox measurements.
OFF – Source power is kept ON at all times, including during trigger Hold
o
mode and between single sweep acquisitions. The RF source stays at the stop
frequency until another sweep begins. This provides a constant,
uninterrupted RF output from the RF OUT port. This behavior also provides
the highest trace stability sweep-to-sweep which may only be evident while
using either single sweep or Run/Hold. With Battery Save OFF, the use of a
full battery charge is reduced by approximately 45 minutes.
System Configuration
Options(Licensing)
Show Options
http://www.agilent.com/find/softwarelicense
There is a beep when your settings have been
to restore the factory default settings.
147