 Then choose from the following:

Smoothing

Trace smoothing averages a number of adjacent data points to smooth the peak-
to-peak noise values on a displayed trace. The number of adjacent data points
that are averaged is known as the smoothing
specifying a percentage of the X-axis span.
Trace smoothing does NOT significantly increase measurement time.
Smoothing is used in Group Delay measurements, although it can be used with
any NA format EXCEPT Polar or Smith Chart. Learn more about NA Mode
formats, including Group Delay, on page 40.
When enabled,
How to set Smoothing
 Press BW 2.
 Then
 Then
 Press Enter
Single or Continuous Measure
This setting determines whether the FieldFox measures continuously or only
once each time the
power or to allow you to save or analyze a specific measurement.
This setting can be changed at any time without affecting calibration accuracy.
How to set Single or Continuous
 Press Sweep 3.
 Then choose one of the following:
You can also use Run / Hold +/- to toggle between Single and Continuous.
Resolution (Number of Data Points)
Data points are individual measurements that are made and plotted across the X-
axis to form a trace. Select more data points to increase measurement resolution.
However, more data points also takes more time to complete an entire
measurement sweep.
When the Resolution is changed after a calibration is performed, the cal becomes
interpolated. Learn more on page 54.
44
Smo
appears on the FieldFox screen.
Smoothing ON OFF
Sm. Aperture
and enter a value between 0 and 25 (percent) using the
numeric keypad.
Single
Automatically sets Continuous OFF and causes FieldFox to make
Single
o
ONE measurement, then hold for the next Single key press. When a data
trace is displayed, the entire trace is measured, then holds. The
annotation changes to an arrow
Continuous
Makes continuous measurements. This is the typical setting
o
when battery power is not critical.
300 Hz | 1 kHz | 3 kHz | 10 kHz | 30 kHz
aperture.
button is pressed. Use Single to conserve battery
-->
while the measurement occurs.
Aperture is set by
Hold
N9912A FieldFox User's Guide