Field Strength Measurements
Trace 1 - Corrected trace with antenna factor. (Antenna = ON, Apply Corr = ON)
Trace 2
Trace 4
To see the combined Antenna and Cable correction curve, set
state (learn how on page 78) with ALL relevant corrections set to ON.
 Learn how to set Y-Axis Units on page 63.
 Use a Band Power marker to measure total power over a range of frequencies.
How to select correction for Field Strength measurements
The Antenna and Cable correction data survives a Mode Preset and Preset.
All Correction ON/OFF states survive a Mode Preset, but NOT a Preset.
 Press Scale/Amptd .
 Then
 Then
 Then choose from the following:
Using the Antenna/Cable Editor
Two (appended) pages of Antenna Editor
The Antenna Editor and the Cable Editor menus are identical. Both tables
include header information, and a Frequency/Value table.
SA (Spectrum Analyzer) Mode
- (View) Uncorrected trace (Apply Corr = OFF)
- (View) Current correction factors are automatically stored in Trace 4.
Learn how on page 130.
More
Corrections
Apply Corrections ON OFF
o
Antenna ON OFF
Turn ON and OFF Antenna correction.
o
Edit/Save/Recall Antennas
o
below.
Turn ON and OFF Cable correction.
Cable ON OFF
o
Edit/Save/Recall Cables
Turn ON and OFF correction for all settings
Recall or edit an Antenna Correction table. See
Recall or edit a Cable Correction table. See below.
Trace 4
to View