Factory Settings - ABB AX410 User Manual

Single and dual input analyzers for low level conductivity
Hide thumbs Also See for AX410:
Table of Contents

Advertisement

Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
Sensor A Terminal Numbers
Sensor B Terminal Numbers
Fig. 7.1 Analyzer Terminal Links and Decade Resistance Box Connections

7.3 Factory Settings

Section 7.3, Page 63
FACTORY SETTINGS
Use the Menu
Key to scroll
through
the Menus
To
OPERATING PAGE
Cal. Output 2
O2: Adjust 4mA
O2: Adjust 20mA
Key
Available only if the option board is fitted
and analog features enabled – see page 67
Dual conductivity only
Fig. 7.2 Overall Factory Settings Chart
Note. The Option Board page is displayed only if the option board is fitted.
62
Temperature
Simulator
B9
B10
B1
B2
Terminal link
Use the Sidescroll Key to scroll through the Pages within each Menu
Factory Set Code
Cal. Sensor A
Calibrate I/P ?
A:Res.Zero (O/C)
A:Res.Span (2K0)
Use the Downscroll
Key to scroll through
A:Res.Zero (O/C)
the Frames
A:Res.Span(20R)
within each Page
A: Self Checking
A:T.Zero
A:T.Span
A:T.Zero
A:T.Span
Cal. Output 3
Cal. Output 4
O3: Adjust 4mA
O4: Adjust 4mA
O3: Adjust 20mA
O4: Adjust 20mA
B11 B12
B13
B14
B3
B4
B5
B6
Cal. Sensor B
Calibrate I/P ?
B:Res.Zero (O/C)
B:Res.Span (2K0)
B:Res.Zero (O/C)
B:Res.Span(20R)
B: Self Checking
(100R)
B:T.Zero
(100R)
(150R)
B:T.Span
(150R)
(1K0)
B:T.Zero
(1K0)
(1K5)
B:T.Span
(1K5)
Option Board
7 CALIBRATION
Conductivity
Cell Simulator
Earth
Cal. Output 1
O1: Adjust 4mA
O1: Adjust 20mA
Alter Fact.Code
IM/AX4CO
Issue 11

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ax411Ax413Ax416Ax418Ax456Ax450 ... Show all

Table of Contents