Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
-----
Cal. Output 1
16000
O1: Adjust 4mA
7200
O1: Adjust 20mA
Cal. Output 1
-----
Cal. Output 2
7200
O1: Adjust 20mA
Cal. Output 2
66
Calibrate Output 1
Note. When adjusting the 4 and 20mA output values, the display reading is
unimportant and is used only to indicate that the output is changing when the
and
keys are pressed.
See below.
Cal. Output 2
Adjust 4mA
Use the
Note. The analog output range selected in
affect the reading.
Adjust 20mA
Use the
Note. The analog output range selected in
affect the reading.
See below.
Cal. Output 2
Operating Page - see section 2.3, page 6.
Conductivity
Dual Cond.
Calibrate Output 2
Note. Output 2 calibration is identical to Output 1 calibration.
Option board fitted and analog features enabled - continued on next page.
Cal. Output 3
Option board fitted, additional features disabled - continued on next page.
Option Board
Option board not fitted - continued on next page.
Alter Fact. Code
Option board fitted and analog features enabled - continued on next page.
Cal. Output 3
Option board fitted, additional features disabled - continued on next page.
Option Board
Option board not fitted - continued on next page.
Alter Fact. Code
Conductivity
Operating Page - see section 2.3, page 6.
Dual Cond.
and
keys to set the milliammeter reading to 4mA.
and
keys to set the milliammeter reading to 20mA.
7 CALIBRATION
(Section 5.5) does not
Configure Outputs
(Section 5.5) does not
Configure Outputs
IM/AX4CO
Issue 11