Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456

6.4.2 Connections

Earth (Ground)
Stud on Case
(see Fig. 6.8.)
Fig. 6.9 Connections, Wall-/Pipe-mount Analyzer
Note.
When connecting non-metal conductivity cells or metal conductivity cells that are isolated from earth (ground), e.g.
mounted in plastic pipework, link Terminal B14 (and Terminal B6 if dual input analyzer) to the earth (ground) stud on the
analyzer case - see Fig. 6.8.
When connecting earthed (grounded) metal conductivity cells, ensure that the cell earth (ground) and the analyzer earth
(ground) stud are at the same potential.
IM/AX4CO
Issue 11
Terminal Block C
(Option Board)
Terminal Block B
Sensor B
Terminal block A
B1
B2
B3
B4
B5
B6
B7
B8
* When a 2-wire Pt100, Pt1000 or 'balco 3k' temperature compensator is fitted.
Refer also to Section 6.6 for ABB UK and US Sensor Systems Connections
Terminal Block B
Temperature Compensator
Connections
Before making any electrical connections,
see Warnings on page 53
Sensor A
Conductivity Cell Connections
Temperature Compensator Common.
B9
Link B9 to B10 (and B1 to B2 if Dual Input Analyzer) *
B10
Temperature Compensator Third Lead
B11
Temperature Compensator
B12
B13
Cell (Cell Electrode)
B14
Cell (Earth Electrode) – see Note below
B15
B16
6 INSTALLATION
Temperature Compensator
Connections
Screen
Not Used
Not Used
57