Advertisement

Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
0. 0
uS/cm
0. 0
Calib
A:Res.Zero (O/C)
500. 0
uS/cm
500. 0
Calib
A:Res.Span (2K0)
0
uS/cm
0
Calib
A:Res.Zero (O/C)
64
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Note. The upper 7-segment display shows the measured conductivity. Once the signal
is within range the lower 7-segment display shows the same value and
displayed to indicate that calibration is in progress.
Resistance Span (2k0)
Set the cell simulator to 2kΩ.
The display advances automatically to the next step once a stable and valid value is
recorded.
Resistance Zero (Open Circuit)
Open circuit the cell simulator.
The display advances automatically to the next step once a stable and valid value is
recorded.
Continued on next page.
A:Res.Span(20R)
7 CALIBRATION
is
Calib
IM/AX4CO
Issue 11