A.3.4 Monitoring on Solid Alkaline Treated Systems
Generally, boiler waters treated with solid alkaline chemicals, for
example, sodium hydroxide, have relatively high conductivities.
The dual input conductivity analyzer, in conjunction with a cation
resin column, can be used to indicate sample pH. If the sample
also contains salts (e.g. sodium chloride), the after-cation
conductivity reading reflects the acid conductivity released by
the salts; the reading is typically three times higher than normal
owing to the acid. Hence to derive the concentration and pH of
the alkaline agent, one third of the after-cation conductivity
increase must be subtracted from the before-column reading. In
addition, a factor must be applied for the molar conductivity
change of the alkaline agent. The analyzer software applies the
following equation:
log(BC –
Inferred pH = 11 +
Where: BC = the before column reading
AC = the after column reading
F
= molar conductivity change for the alkaline
agent (243μS cm
hydroxide*)
The maximum after-cation conductivity value is programmable
between 1.00 and 100.0μS cm
After-cation values above this level generate an
error message and before-cation values above 100.0μS
HIGH
cm
-1
generate a
BEFORE CAT. HIGH
pH range is 7 to 11pH; values above 11pH generate an
error message. Refer to Section 8 for description of error
invalid
messages.
The inferred pH feature can be used on solid alkali treated
systems only in the following circumstances:
1. On steam raising plant.
2. For boiler chemical treatment such as sodium hydroxide.
In this instance,
A: Temp. Comp.
must be set to
B: Temp. Comp.
page 21.
3. Where the after-cation conductivity value is less than
100.0μS cm
-1
.
Note. Inferred pH measurement on solid alkaline treated
systems is inappropriate to samples containing sodium
phosphate, ammonia or morpholine.
* Refer to Appendix to VGB Guideline VGB-R 450 L.
IM/AX4CO
Issue 11
1
/
AC)
3
F
-1
per mmol/l for sodium
-1
dependent on local conditions.
AFTER CAT.
error message. The inferred
must be set to
NaOH
- see section 5.3,
ACID
Infr. pH
and
Appendix A
77