8.3 Checking the Temperature Input

Check the analyzer responds to a temperature input.
Disconnect the Pt100/Pt1000 leads and connect a suitable
resistance box directly to the analyzer inputs - see section 7.2,
page 61. Check the analyzer displays the correct values as set
on the resistance box - see Table 8.3.
Temperature
ºC
ºF
0
32
10
50
20
68
25
77
30
86
40
104
50
122
60
140
70
158
80
176
90
194
100
212
130.5
267
Table 8.3 Temperature Readings for Resistance Inputs
Incorrect readings usually indicate an electrical calibration
problem. Re-calibrate the analyzer as detailed in Section 7.3.
IM/AX4CO
Issue 11
Input Resistance (Ω)
Pt100
Pt1000
100.00
1000.0
103.90
1039.0
107.79
1077.9
109.73
1097.3
111.67
1116.7
115.54
1155.4
119.40
1194.0
123.24
1232.4
127.07
1270.7
130.89
1308.9
134.70
1347.0
138.50
1385.0
150.00
1500.0
8 SIMPLE FAULT FINDING