Single and Dual Input Analyzers for Low Level Conductivity
AX410, AX411, AX413, AX416, AX418, AX450, AX455 & AX456
Hi TC
Lo TC
-----
Off
A: T.Comp Range
Lo TC
A: Temp.Comp
Hi TC
A: Temp.Comp
Off
A: Temp.Sensor
IM/AX4CO
Issue 11
Temperature Compensation Range
Select a temperature compensation range to suit the sample temperature:
Off
-
Raw conductivity measurement without temperature compensation.
Examples
- Water for injection (WFI) for US Pharmacopoeia (USP) applications.
- Purified water for USP applications.
Lo TC
-
Temperature compensation for sample temperatures in the range 0 to
100ºC (0 to 212ºF). This setting is suitable for the majority of applications.
Hi TC
-
Temperature compensation for sample temperatures in the range 0 to
200ºC (0 to 392ºF). This setting is used only for special high temperature
applications.