Removing And Replacing The A40 Floppy Disk Drive - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the A40 Floppy Disk Drive

Removing and Replacing the A40 Floppy Disk Drive
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-25
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. Remove the front panel assembly. Refer to
Assembly" on page
4. Remove the A13 (Option 080) frequency offset receiver assembly, if present. Refer to
"Removing and Replacing the A13 Frequency Offset Receiver Assembly" on page
5. Remove the A12 source 20. Refer to
page
7-22.
6. Disconnect the ribbon cable (item
7. Lift the tab at the rear of the A40 floppy disk drive enclosure and remove the rear
enclosure cover (item
8. Slide the A40 floppy disk drive out of the front of the enclosure. The A40 floppy disk
drive fits snugly. You may have to push the drive from the back to remove it from the
analyzer.
Replacement Procedure
1. Reverse the order of the removal procedure using the existing ribbon cable (item
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-54
and
Figure 7-26
for this procedure.
7-10.
"Removing and Replacing the A12 Source 20" on
) from the A14 system motherboard.
).
Table 7-2 on page
PNA Series Microwave Network Analyzers
"Removing the Covers" on page
"Removing and Replacing the Front Panel
7-66.
Service Guide E8361-90001
E8361A
7-6.
7-24.
).