Removing And Replacing The A12 Source 20 - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Repair and Replacement Procedures

Removing and Replacing the A12 Source 20

Removing and Replacing the A12 Source 20
Tools Required
• T-10 TORX driver (set to 9 in-lb)
• T-20 TORX driver (set to 21 in-lb)
• 5/16-inch open-end torque wrench (set to 10 in-lb)
• ESD grounding wrist strap
Removal Procedure
Refer to
Figure 7-9
for this procedure.
1. Disconnect the power cord.
2. Remove the outer and inner covers. Refer to
3. Remove the A13 (Option 080) frequency offset receiver assembly, if present. Refer to
"Removing and Replacing the A13 Frequency Offset Receiver Assembly" on page
4. With a 5/16-inch torque wrench, disconnect the semirigid cable (item
bottom of the A12 source 20.
CAUTION
To avoid damage to the semirigid cable (item
the bottom of the analyzer to move it away while sliding out the
A12 source 20. The cable has some spring to it and the RF pin could be
damaged by the A12 source 20.
5. Disengage the A12 source 20 from the A11 connector and slide out of the analyzer.
Observe the
CAUTION
Replacement Procedure
1. Reverse the order of the removal procedure.
2. Perform the post-repair adjustments, verifications, and performance tests that pertain
to this removal procedure. Refer to
7-22
above.
Table 7-2 on page
PNA Series Microwave Network Analyzers
"Removing the Covers" on page
), pull down on the cable from
7-66.
Service Guide E8361-90001
E8361A
7-6.
7-24.
) from the

Advertisement

Table of Contents
loading

Table of Contents