Signal Separation Group Operation - Agilent Technologies PNA Series Service Manual

Microwave network analyzer
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

Theory of Operation

Signal Separation Group Operation

Signal Separation Group Operation
The signal separation group divides the source signal into a reference path and a test path.
Refer to
Figure 5-3 on page
• The reference signal is transmitted to the receiver group.
• The test signal is transmitted through—and reflected from—the device under test
(DUT) and then is transmitted to the receiver group.
• Control lines to this group are routed from the A16 test set motherboard.
In this section, the following assemblies are described:
A22 Switch
A23 and A24 Source Multiplier/Amplifier 70 (SOMA 70)
A25 and A26 Test Port Couplers
• Front Panel Jumpers
Figure 5-3 Signal Separation Group, Standard Analyzer
5-16
5-16.
(Option 014 Configurable Test
PNA Series Microwave Network Analyzers
Set)